OHMEDA CARE PLUS INFANT INCUBATOR

$1,200.00
Ex Tax: $1,200.00
Availability: In Stock
Brand: DATEX-OHMEDA
Product Code: OHMEDA CARE PLUS INFANT INCUBATOR
Product viewed: 91
Qty:     

DETAILS
Integrates microprocessor-based controls with a modular design for efficient patient care
 is an advanced incubator that provides superior thermoregulation with an integrated controller based on a microprocessor. The optional active double wall construction creates a precise and consistent environment for the newborn.

Features
Microprocessor-based control system for precise thermoregulation.
Self-test functions during power up and during normal operation.
Patient control mode (Servo) and air control mode (Manual)
Continuous positioning from 0 ° to 12 °.
Optional construction of double active wall.

Dimensions
Incubator with lifting base
Height: 54 - 62 in (136.2 - 157.5 cm)
Base of elevation: 35 - 43 in (88.9 - 109.2 cm) mattress on the floor
Depth: 28 in (71.1 cm)
Width: 33 in (83.8 cm)
Weight: 195 pounds (89 kg)
Incubator with stationary pedestal
Height: 53 in (135.9 cm)
Depth: 28 in (71.1 cm)
Width: 33 in (83.8 cm)
Weight: 180 pounds (80 kg)
Packages of drawers for lifting bases or stationary pedestals
In general: 11 x 11.5 x 18.25 in (28 x 29 x 46 cm)
Individual deep drawer: 10 x 9 x 16.5 in (25 x 23 x 42 cm)
Double drawer: 4.5 x 9 x 16.5 in (11 x 23 x 42 cm)
Incubator with cabinet
Height: 55 in (141 cm)
Depth: 25.5 in (64.6 cm)
Width: 35 in (88.9 cm)
Weight: 185 pounds (84 kg)
Characteristics of the incubator
Wheels: 5 inches in diameter, 2 with lock, 2 without lock
Mattress: 13.7 x 25.6 in (34.8 x 65 cm)
From mattress to hood: 16 in (41 cm)
Door height: 11 in (28 cm)
Positions: continuous from 0 ° to 12 °.
Port holes: 6
Pipe ports: 6
Electric
120 VAC 60 Hz
Models: 115 V ± 10% 5.7 amps.
Meets the requirements of IEC 601.1, UL 544 and CSA 22.2 No. 125
Rated power consumption: 450 watts at the maximum power of the heater
Line voltage compensation: the input voltage is monitored and the heat output adjusted to compensate for line voltage variations
Leakage currents: less than 100 micro amps in 120 V units
Patient probe leakage: less than 50 micro amps in 120 V units
Ambient
Operating temperature range: 68 ° - 86 ° F (20 ° - 30 ° C)
Storage temperature range: 13 ° - 140 ° F (-25 ° - 60 ° C)
Operating humidity range: 0 to 90%
System controls
Microprocessor-based control system: self-test functions are performed during power-up and during normal operation
Patient control mode (servo): 95 ° - 98.6 ° F (35 ° - 37 ° C)
Air control mode (manual): 73.4 ° - 98.6 ° F (20 ° - 37 ° C)
Up to 102.2 ° F (39 ° C) with control panel override switch
Air speed: average <10 cm / sec with interior wall
Temperature variability: less than ± 0.2 ° F (0.1 ° C)